The instrument page holds a gauge's details and its calibration history. Here you record each calibration with its as-found and as-left results; an out-of-tolerance result raises an NCR or quarantines the instrument.
Add an instrument
- Open QualityCalibration and select New instrument.
- Enter the Code, Name and Type, the Serial number, Manufacturer, Model, Location and Owner.
- Under Metrology, enter the Measuring range, Resolution, Tolerance ±, the Calibration provider and the Calibration interval (days), and the Next due date if known.
- Select Create instrument. The message Instrument … created appears.
Record a calibration
- Open the instrument and select Record calibration (needs Permission:
quality.calibration.manage). - Enter the Date, Performed by, Calibration provider and Certificate number.
- Under Test points, select Add test point for each point checked: the Test point, Reference value, As found and As left readings and the Tolerance ± (the instrument's tolerance is suggested). Points are judged against the tolerance. Without test points, give the As found and As left results directly.
- Tick The instrument was adjusted if it was.
- Check Next due (empty: date + the interval) and add Notes.
- Leave Raise an NCR when found out of tolerance on. The NCR lists the inspections measured with this instrument since its last calibration, so you can judge whether they are still valid.
- Select Save calibration.
| Result | What happens |
|---|---|
| As found and as left in tolerance | Calibration … recorded; next due date moves on. |
| As found out of tolerance | Calibration … recorded — out of tolerance, NCR-… raised (source Calibration). |
| As left out of tolerance | The instrument is Quarantined: Recalibrate it before use. |
Calibration history lists every calibration with its number (CAL-2026-00001), date, As found, As left, Adjusted, Certificate, NCR and a PDF certificate.
Instrument statuses
| Status | Meaning |
|---|---|
| Active | In use. |
| Quarantined | Left out of tolerance; recalibrate before use. |
| Out of service | Temporarily not used (set in the Status field). |
| Retired | Removed with Retire instrument; leaves the board and can no longer be calibrated or used in plans. |